Outline

Recently, demands for measuring near-infrared spectral region have increased, however, especially in FPD and lighting market, there is not handy and high accuracy measuring instrument.

Under these circumstances, we have developed NIR Spectroradiometer SR-NIR based on technology of our famous SR-3AR. And, you can measure Spectral radiance from 380nm to 1030nm using by both other SR series and SR-NIR.

Features

  • Measurable ultra weak optical signal of near-infrared spectral region from flat panel display
     
  • High accuracy measuring of near-infrared spectral region.
     
  • Measurable near-infrared spectral form 380nm to 1030nm using by both other SR series and SR-NIR.

Applications

  • For measuring near infra-infrared spectral region from FPD such as LCD and OLED.
  • For measuring Nr, Ar line spectra
  • For measuring transmission characteristics of film sheet in near-infrared spectral region
  • Nir infra-infrared spectral region from other light sourcebr />

Specifications

Photo detector Electronically cooled back incidence type CCD
Dispersing element Diffraction grating
Measurement angle 2°/1°/0.2°/0.1° selectable (Motor moved)
Minimum measurement area 2° : 10.0,   1° : 4.99,   0.2° : 1.00,   0.1° : 0.50 (mmφ)
Measurement distance 350mm - ∞ (distance from metallic tip of objective lens)
Spectral bandwidth 6 - 8nm (half bandwidth)
Wavelength range 600 - 1030nm
Wavelength resolution 1nm
Measurement mode Auto/Manual (integral time/frequency)
Measurement object Spectral radiance : W・sr-1・m-2・nm-1
Calculation function Spectral radiance 600 - 1030nm
Measurement range(*1) 2.0° : 0.5 - 3,000,
1.0° : 1 - 9,000
0.2° : 20 - 70,000,
0.1° : 100 - 300,000
Repeatability(*2) ±2% or less
Polarization error Spectral radiance 5% or less
Calibration standard Topcon standard lamp (Standard illuminance A, 23℃±3℃, 65%RH or less)
Measurement time(*3) About 1 - 31 sec
Interface USB2.0 / RS-232C
Power supply Dedicated AC adapter AC100V-240V, 50/60Hz
Power consumption about 34W
Use conditions Temperature: 5 - 30℃, Humidity: 80%RH or less (no condensation)
Outer dimensions (W×D×H) 150×406×239mm
Weight 5.5kg

(*1) :This product is no brightness measurement, so this is the reference data for standard illuminance A

(*2) : 600 - 1030nm, For TOPCON standard illuminance

(*3) : Except communication time with PC.

Measurement area (Diameter φ)

(mmφ)
Measurement angle Measurement distance(*1)
350mm 500mm 800mm 1000mm 2000mm
10 15.1 25.4 32.2 66.4
4.99 7.55 12.7 16.1 33.2
0.2° 1.00 1.51 2.54 3.22 6.64
0.1° 0.50 0.76 1.27 1.61 3.32

(*1) : The measurement distance is the distance from the tip of the metal fixture on the instrument of the objective lens.

Available Accessories

  • Attachment Lens
: AL-6/AL-11/AL-12
  • Reference White Board
: WS-3
  • Fiber probe
: FP-3P
  • Tripod 5N
: Tripod5N
  • Tripod
: Tripod-SR
  • Fine Control Table
: S-4
  • CCD Adapter
: IA-2