Topcon 產品諮詢專線:04-27080265 / 劉小姐 |
Outline
Recently, luminance and chromaticity management of FDD, Automotive instruments, Fluorescent substance and optical device has been increased. Especially, the needs for more measured weak optical signal and faster instrument are growing. We achieve with high chromaticity and measuring speed at BM-5AC, compared with current model BM-5AS. |
Features
|
Applications
|
Specifications
Photo detector | Photomultiplier tube |
---|---|
Optical system | Objective lens : f=80mm F2.5, Eyepiece lens : View field 5° |
Spectral sensitivity | Similar to CIE1931 color matching function |
Measurement angle | 3°/2°/1°/0.2°/0.1° *selectable |
Measurement distance | 350mm - ∞ (distance from metallic tip of objective lens) |
Measurement range | Auto/Manual 5-range |
Function(*1) |
|
Measurable range(*2) | 0.00005 - 1,200,000 cd /m² |
Accuracy in luminance(*3) | ±4% or less |
Accuracy in chromaticity | (1) dx, dy : ±0.005 or less(*3) (2) dx, dy : ±0.008 or less(O-55, Y-48, A-73B, IRA-05, T-44,R-61, B-46, V-44, G-54) (*4) (3) dx, dy : ±0.005 or less (*6) |
Repeatability in luminance(*6) | 2.0% : 0.005-0.025cd/m² 0.8% : 0.025 cd/m² or more |
Repeatability in chromaticity(*7) | 0.003 |
Measurement time | About 2sec |
Response speed |
|
Display | Dot-matrix 20caracters x 4rows with backlight |
Interface | RS-232C、 USB |
Power supply | Dedicated AC adapter AC100V-240V, 50/60Hz |
Power consumption | about 20W |
Use conditions | Temperature : 0-40℃, Humidity : 85% RH or less (no condensation) |
Outer dimensions (W×D×H) | 154×355×212mm |
Weight | 3.6kg |
(*1) : ±⊿ : difference of standard value (*2) : For standard illuminance A (*3) : For standard illuminance A, Auto Rage (*4) : with combination of standard illuminant and color glass filter O-55,Y-48,A-73B,IRA-05,T-44,R-61,B-46,V-44,G-54) (*5) : Our specification LED panel of RGBW (*6) : For standard illuminance A, measuring angle 3°, 2σ, Single mode, Auto Range (*7) : For standard illuminance A, 2σ, Single mode, Auto Range (*8) : Single measurement mode |
Measurement area (mm Φ)
(mmφ) |
Measurement angle | Measurement distance (m) (*1) | ||||
---|---|---|---|---|---|
350mm | 500mm | 1000mm | 5000mm | 10000mm | |
3° | 15.0 | 23.1 | 49.2 | 255 | 510 |
2° | 10 | 15.4 | 32.8 | 169 | 341 |
1° | 5 | 7.7 | 16.4 | 85 | 170 |
0.2° | 1 | 1.5 | 3.3 | 17 | 34 |
0.1° | 0.5 | 0.8 | 1.6 | 8 | 17 |
(*1) The measurement distance is the distance from the tip of the metal fixture on the instrument of the objective lens. |
Luminance Display Range
(cd/ m²) |
Measurement angle | Range-1 | Range-2 | Range-3 | Range-4 | Range-5 |
---|---|---|---|---|---|
3° | 0.00005-0.15 | 0.0005-1.5 | 0.005-15 | 0.05-150 | 0.5-1,500 |
2° | 0.0001-0.3 | 0.001-3 | 0.01-30 | 0.1-300 | 1-3,000 |
1° | 0.0004-1.2 | 0.004-12 | 0.04-120 | 0.4-1,200 | 4-12,000 |
0.2° | 0.01-30 | 0.1-300 | 0.4-1,200 | 10-30,000 | 100-300,000 |
0.1° | 0.04-120 | 0.4-1,200 | 4-120,00 | 100-300,000 | 400-1,200,000 |
Available Accessories
|